About Assure

Products center

Solution

News center

Contact us

400-853-5577

Products

Deepen the industry and seek common development

Chronic Disease Series


15 Sec.     10μl       1000memory

Specification:


Specimen Specimen Volume Test Time Measurement Range
Whole blood (capillary and venous human whole blood) 10 μL ≤ 15 seconds 4.5-25.6 g/dL, 45-256 g/L, 2.8-15.9 mmol/L
Memory Meter Storage Conditions System Operating Conditions Battery Life
1000 records 0 - 50°C; ≤ 90% RH 2 - 40°C; 90% RH; height 1,000 tests


5 Sec.      1 μL        200 memory

Specification:

Sample Sample size Measuring time Measurement Range
Fresh capillary or venous whole blood Approximate 1 microlitre 5 seconds 20~600mg/dL(1.1-33.3mmol/L)
Memory Storage Conditions System Operating Conditions Battery Life
Stores up to 200 results with date and time Strip: 2~30°C     Meter: -40~50°C; ≤90%RH 8~37°C 6 months or approximately 1,000 tests



2 Min.         35 µL        500 memory

Specification:


Specimen Specimen Volume Test Time Measurement Range
Capillary blood 35 µL ≤ 2 min TC: 2.59-12.93 mmol/L(100-500 mg/dL ,1mmol/L=38.66 mg/dL)
HDL: 0.39-2.59 mmol/L (15-100 mg/dL, 1mmol/L=38.66 mg/dL)
TG: 0.51-7.34mmol/L (45-650mg/dL, 1mmol/L=88.6 mg/dL)
Memory Storage Conditions System Operating Conditions Battery Life
500 records Meter: 0- 45°C; 90% RH Test Device: 2 - 30°C; 90% RH Check Device: 2 - 30°C; 90% RH 10 - 35°C; ≤ 80% RH; height 1,000 tests




Sample Sample size Measuring time Measurement range
Fresh capillary whole blood Blood Glucose:
Approximate 1 microlitre
Uric acid: Approximate 2 microlitre

Blood Glucose:5 seconds Uric acid: 15 seconds Glucose: 1.1-33.3mmol/L (20~600mg/dL)
Uric acid: 181—1188umol/L
3-20mg/dL
Memory Storage Conditions System operating temperature Battery life
Stores 300 results of blood glucose testing and 100 results of uric acid testing Strip: 2-30°C (35.6-86°F)
Meter: -20~50°C(-4~122°F); 90%RH

10~40°C (50-104°F) 6 months or approximately 1,000 tests

Sample Sample size Measuring time Measurement range
Fresh Capillary or venous whole blood Blood Glucose: 0.8 microlitre
β-Ketone: 1.2 microlitre
Blood Glucose: 5 seconds
β-Ketone: 8 seconds

Glucose: 20~600mg/dL(1.1-33.3mmol/L)
β-Ketone: 0~8.0mmol/L
Memory Storage Conditions Operating temperature Battery life
Stores 300 results of blood glucose testing and 100 results of β-ketone testing. Strip: 2~30°C (36~86°F) Meter: -5~45°C; £90%RH 10~40°C 100~150 hours or approximately 1,000 tests
Return